How To Create The Right Flow Cytometry Antibody Panel Every Time

Sudoku puzzles seem to be all the rage.

I see it in coffeehouses, at the airport, even in doctors offices. Everyone is trying to work out how to fit the numbers into the grids so that everything adds up properly.

Designing polychromatic flow cytometry panels is much like the Sudoku puzzle.

In this case, the grid is composed of the antigens on one side, and the cytometer detectors on the other.

The goal is to fill in the grid correctly.  

Instead of adding up to 45, like in Sudoku, the flow cytometrist is trying to optimize the ability to make a sensitive measurement to answer the biological question the researcher has set out to answer.

Solving the Polychromatic Sudoku puzzle, so to speak, is easy if you follow a few simple steps.

Step 1:  Establish a biological hypothesis.

Everything starts with understanding what the biological hypothesis for the experiments to be performed. This will dictate what populations need to be identified, and what information needs to be extracted from the data.

As you proceed, rank your antibodies based on cellular expression level and importance in answering the biological hypothesis.

For example, CD3 is a highly expressed antigen on T-cells and is important in making primary gating decisions, while CD86 is a dimly expressed (or emergent) marker on cells undergoing activation and may be critical to answer the biological hypothesis.

Step 2:  Research your fluorochromes.

Fluorochrome brightness can be measured and the different fluorochromes ranked against each other.

Use a chart like this one from BioLegend where fluorochromes are ranked from brightest (5) to most dim (1).

Based on the brightness of fluorochromes and the expression density of the antigen on the cells, we want to pair highly expressed antigens (like CD3) with dimmer fluorochromes (1-2), while lower expression antigens are paired with brighter fluorochromes (4,5).

Step 3:  Know the instrument you’re using.

Aim to understand your flow cytometer before designing your antibody panel. Pay particular attention to where the most sensitive measurements can be made on the system. Everything doesn’t boil down to just fluorescence intensity.

Sometimes, it is better to use a less bright fluorochrome if… if the channel does not receive a lot of error.

Below is an example Excel analysis of the spillover of fluorochromes into different detectors on a 4-laser instrument (405, 488. 532. 633).  Summing across results in the error that that a given detector receives from the flurorchromes in the panel. Summing down the columns results in the amount of error a given fluorochrome contributes to the panel in question.

Flow Cytometry Antibody Panel Design

These data are based on the calculations found in Nguyen et al. (2013) Cytometry A 83A:306-315

Using the above analysis, one can make an informed decision where the best fluorochrome choices are to make sensitive measurements.

Step 4:  Use a panel building program.  

It becomes critical to find all the antigen-fluorochrome pairs that are available.

While it is possible to search through catalogues and use Google, there are packages out there that can be used to help this process, for example:

Chromocyte – a web-based system panel building system and a resource website.

Fluorofinder – a new resource with a web-based interface for panel building.

With these tools, the searching for reagents becomes easy, leaving time to build the panels using the theory described above.

Step 5: Optimize your panel. 

After all is said and done, the panel must be optimized and validated.  Proper antibody titration, the correct voltages, optimized staining protocols to minimize non-specific binding and such are all part of the long, but very critical process to ensure that the panel works appropriately.

Step 6:  Bring on the OMIP.

OMIP is short for Optimized Multicolor Immunophenotyping Panel, which is a peer-reviewed, optimized flow panel. The beauty of the OMIP is the work is already done for you, including the antigens to be used, the fluorochromes, and the analysis template.

There are currently over 23 OMIPs, all of which can save you the headache of building a panel from scratch.

Of course, there are still an uncountable number of antibody panels for which no OMIP exists. If you’re designing a new antibody panel, or are stuck after several attempts to optimize a existing panel, get expert help. Don’t keep wasting time and resources on an incorrectly designed panel.

To learn more about automatic compensation and to get access to all of our advanced materials including 20 training videos, presentations, workbooks, and private group membership, get on Mastery Class wait list.

Join Expert Cytometry's Mastery Class

ABOUT TIM BUSHNELL, PHD

Tim Bushnell holds a PhD in Biology from the Rensselaer Polytechnic Institute. He is a co-founder of—and didactic mind behind—ExCyte, the world’s leading flow cytometry training company, which organization boasts a veritable library of in-the-lab resources on sequencing, microscopy, and related topics in the life sciences.

Tim Bushnell, PhD

Similar Articles

Fickle Markers: Solutions For Antibody Binding Specificity Challenges

Fickle Markers: Solutions For Antibody Binding Specificity Challenges

By: Tim Bushnell, PhD

Reproducibility has been an ongoing, and important, concept in the sciences for years.  In the area of biomedical research, the alarm was sounded by several papers published in the early 2010’s.  Authors like Begley and Ellis, Prinz and coworkers, and Vasilevsky and colleagues, among others reported an alarming trend in the reproducibility of pre-clinical data.  These reports indicated between 50% to almost 90% of published pre-clinical data were not reproducible.  This was further highlighted in the article by Freedman and coworkers, who tried to identify and quantify the different sources of error that could be causing this crisis.  Figure 1,…

The Fluorochrome Less Excited: How To Build A Flow Cytometry Antibody Panel

The Fluorochrome Less Excited: How To Build A Flow Cytometry Antibody Panel

By: Tim Bushnell, PhD

Fluorochrome, antibodies and detectors are important. The journey of a thousand cells starts with a good fluorescent panel. The polychromatic panel is the combination of antibodies and fluorochromes. These will be used during the experiment to answer the biological question of interest. When you only need a few targets, the creation of the panel is relatively straightforward. It’s only when you start to get into more complex panels with multiple fluorochromes that overlap in excitation and emission gets more interesting.  FLUOROCHROMES Both full spectrum and traditional fluorescent flow cytometry rely on measuring the emission of the fluorochromes that are attached…

Flow Cytometry Year in Review: Key Changes To Know

Flow Cytometry Year in Review: Key Changes To Know

By: Meerambika Mishra

Here we are, at the end of an eventful year 2021. But with the promise of a new year 2022 to come. It has been a long year, filled with ups and downs. It is always good to reflect on the past year as we move to the future.  In Memoriam Sir Isaac Newton wrote “If I have seen further, it is by standing upon the shoulders of giants.” In the past year, we have lost some giants of our field including Zbigniew Darzynkiwicz, who contributed much in the areas of cell cycle analysis and apoptosis. Howard Shapiro, known for…

What Star Trek Taught Me About Flow Cytometry

What Star Trek Taught Me About Flow Cytometry

By: Tim Bushnell, PhD

It is no secret that I am a very big fan of the Star Trek franchise. There are many good episodes and lessons explored in the 813+ episodes, 12 movies (and counting). Don’t worry, this blog is not going to review all 813, or even 5 of them. Instead, some of the lessons I have taken away from the show that have applicability to science and flow cytometry.  “Darmok and Jalad at Tanagra.”  (ST:TNG season 5, episode 2) This is probably one of my favorite episodes, which involves Picard and an alien trying to establish a common ground and learn…

5 Flow Cytometry Strategies That Sun Tzu Taught Me

5 Flow Cytometry Strategies That Sun Tzu Taught Me

By: Tim Bushnell, PhD

Sun Tzu was a Chinese general and philosopher. His most famous writing is ‘The Art of War’, and has been studied by generals and CEOs, to glean ideas and strategies to help their missions. I was recently rereading this work and thought to myself if any of Sun Tzu’s lessons could apply to flow cytometry.  So I have identified 5 points that I think lend themselves to thinking about flow cytometry.  “Quickness is the essence of the war.” In flow cytometry, speed is of the essence. The longer the cells are out of their natural environment, the less happy they…

A Basic Guide To Flow Cytometry (3 Foundational Concepts)

A Basic Guide To Flow Cytometry (3 Foundational Concepts)

By: Meerambika Mishra

Mastering foundational concepts are imperative for successfully using any technique or system.  Robert Heinlein introduced the term ‘Grok’  in his novel Stranger in a Strange Land. Ever since then it has made its way into popular culture. To Grok something is to understand it intuitively, fully. As a cytometrist, there are several key concepts that you must grok to be successful in your career. These foundational concepts are the key tools that we use day in and day out to identify and characterize our cells of interest.  Cells Flow cytometry measures biological processes at the whole cell level. To do…

Which Fluorophores To Use For Your Microscopy Experiment

Which Fluorophores To Use For Your Microscopy Experiment

By: Heather Brown-Harding, PhD

Fluorophore selection is important. I have often been asked by my facility users which fluorophore is best suited for their experiments. The answer to this is mostly dependent on whether they are using a widefield microscope with set excitation/emission cubes or a laser based system that lets you select the laser and the emission window. Once you have narrowed down which fluorophores you can excite and collect the correct emission, you can further refine the specific fluorophore that is best for your experiment.  In this blog  we will discuss how to determine what can work with your microscope, and how…

4 No Cost Ways To Improve Your Microscopy Image Quality

4 No Cost Ways To Improve Your Microscopy Image Quality

By: Heather Brown-Harding, PhD

Image quality is critical for accurate and reproducible data. Many people get stuck on the magnification of the objective or on using a confocal instead of a widefield microscope. There are several other factors that affect the image quality such as the numerical aperture of the objective, the signal-to-noise ratio of the system, or the brightness of the sample.  Numerical aperture is the ability of an objective to collect light from a sample, but it contributes to two key formulas that will affect your image quality. The first is the theoretical resolution of the objective. It is expressed with the…

What Is Total Internal Reflection Fluorescence (TIRF) Microscopy & Is It Right For You?

What Is Total Internal Reflection Fluorescence (TIRF) Microscopy & Is It Right For You?

By: Heather Brown-Harding, PhD

TIRF is not as common as other microscopy based techniques due to certain restrictions. We will discuss these restrictions, then analyze why it might be perfect for your experiment.  TIRF relies on an evanescent wave, created through a critical angle of coherent light (i.e. laser) that reaches a refractive index mismatch.  What does it mean in practice?  A high angle laser reflects off the interface of the coverslip and the sample. Although the depth that this wave penetrates is dependent on the wavelength of the light, in practice it is approximately 50-300nm from the coverslip. Therefore, the cell membrane is…

Top Industry Career eBooks

Get the Advanced Microscopy eBook

Get the Advanced Microscopy eBook

Heather Brown-Harding, PhD

Learn the best practices and advanced techniques across the diverse fields of microscopy, including instrumentation, experimental setup, image analysis, figure preparation, and more.

Get The Free Modern Flow Cytometry eBook

Get The Free Modern Flow Cytometry eBook

Tim Bushnell, PhD

Learn the best practices of flow cytometry experimentation, data analysis, figure preparation, antibody panel design, instrumentation and more.

Get The Free 4-10 Compensation eBook

Get The Free 4-10 Compensation eBook

Tim Bushnell, PhD

Advanced 4-10 Color Compensation, Learn strategies for designing advanced antibody compensation panels and how to use your compensation matrix to analyze your experimental data.