5 Flow Cytometry Strategies That Sun Tzu Taught Me

Sun Tzu was a Chinese general and philosopher. His most famous writing is ‘The Art of War’, and has been studied by generals and CEOs, to glean ideas and strategies to help their missions. I was recently rereading this work and thought to myself if any of Sun Tzu’s lessons could apply to flow cytometry.  So I have identified 5 points that I think lend themselves to thinking about flow cytometry. 

“Quickness is the essence of the war.”

In flow cytometry, speed is of the essence. The longer the cells are out of their natural environment, the less happy they are. If you are performing analysis, after isolation, staining time will be dependent on what the target is. Fortunately, for analysis, it is possible to fix the cells and store them till your time on the cytometer. Of course, care must be taken when fixing cells, as fixatives can impact the quality of the fluorochromes. 

Speed is especially important in cell sorting experiments. The table below shows some calculations for how long a given sort would take based on some assumptions. Including the number of events needed for the downstream experiment and the frequency of the target cell.  

Now the question arises, how to decrease this time?  The answer to that question is to look into getting rid of some of the non-target cells using a depletion technique like magnetic beads. Take the example of 100 million cells with a target cell with a frequency of 0.1%, and needing 100K cells for your downstream application. If we sort at 20K events per second, it would take about 83 minutes to sort. However, if we deplete those cells we don’t need (90%), which would take about 30 minutes, we would be down to only 10 million cells, which we can sort in about 10 minutes. Thus the depletion saves us about 40 minutes. 

“Plan for what is difficult while it is easy, do what is great while it is small.”

At the beginning of the experimental design phase, one thing that should be developed is a statistical analysis plan. By establishing this at the beginning of the experiment, you can avoid common errors such as p-hacking and HARKing, among others.  

More importantly, it also helps to determine the size of the experiment based on the power of the experiment. It is important to remember that you need to plan your experiment based on the amount of resources and money you have. This may mean you cannot answer the original question. For example, trying to prove a 5% difference requires 30 samples of each experiment and control. But you only have funds for 15 samples each. Means you can only prove a 10 or 15% difference. 

The other thing to consider is your threshold, the ⍺ value. This is a measure of the change of committing a Type 1 error. While setting this value, it is worth considering the consequences of that false positive. This handy chart can help  you decide. Don’t rely on the generally accepted standard of 0.05. Rather make sure your threshold is set based on the experimental question. 

“If the enemy leaves a door open, you must rush in.”

We have a plan when we start our experiments. We’re trying to discover some new facts about the cells we are studying. However, sometimes the cells may want to show us something else. I know I am ascribing actions to cells that are not true. But we need to be open to the idea that we may find something unexpected and interesting in our experiments. Especially with high-dimensional experiments, we have opportunities to identify populations previously not observed.  Jonathan Irish called these populations ‘cyto incognito’ as they have been overlooked in more traditional analysis.  

Another way to rush in is to take full advantage of both automated analytical tools and multi ‘omics techniques. Either way, be nimble and ready. 

“Opportunities multiply as they are seized.”

Flow cytometry is a powerful and evolving technology. To be successful in using flow cytometry, it is important to learn about the best practices in the design and execution of your experiments. Sure, your lab may have a coffee-stained old notebook from the PI’s time in the lab, and everyone goes to that to find the protocols that the lab uses. However, are those still the best practices? For example, when we had 3- and 4- color analog cytometers, we would perform some form of manual compensation. Unfortunately because of how these systems worked and processed data, it was very easy to overcompensate the data. Many experiments were saved because common colors used – FITC, PE, PerCP and APC, played relatively nice with each other, spectrally speaking. 

Nowadays, with numerous parameters being the norm, it’s critical to perform automated compensation following the best practices, such as discussed in this blog. With the proliferation of full spectrum cytometers, automated unmixing is critical.  

Another practice that needs to be consigned to the dustbins of history is the Isotype control. This blog discusses the issue in greater detail. In general, the isotype control can only show if blocking was successful and should not be used to set positivity. 

There are many opportunities for further education, and you should take full advantage of what is available. Your experiments and data will thank you . 

“To know your enemy, you must become your enemy.”

This quote may seem as an odd one to include, but what is the ‘enemy’ of flow cytometry? That is bad data, a failed experiment. These enemies represent lost opportunities and wasted time and money. To determine how to prevent the failure, understand where these failures occur and how to avoid the trap. 

Planning is the key to knowing your enemy. 

This starts with knowing the tools we can use to prevent our enemy from ruining our experiments.  That is the role of our controls. From quality controls to compensation controls, and everything in between can help you understand where the failure popped up.

The assay development stage is the time to explore what controls are critical for you to identify your target populations. It also helps understand the expected ranges that these populations may have. Ensure to have or inspect the quality control work being done by the people overseeing the instrument. That will let you know the machine is behaving appropriately. In fact, don’t be afraid to add your own QC step into your experiment and track that as well. 

Next, we need compensation controls to properly address spectral spillover. Ensure that we can identify our populations in the presence of all the other fluorochromes. Don’t forget the FMO control as well. It is useful in helping to set the negative/positive boundary. 

 Another useful control is a reference control. This is a standard sample that you run with each experiment that behaves in a predictable way in your assay.  

One other trick I like to do when I am running a polychromatic panel is to have a plot of time vs fluorescence for each laser. Since a blockage on the waste side of the instrument can cause a slowdown of the flow. Prevent the cells from being interrogated by the appropriate laser at the correct time. So you will see a loss of signal as shown below.  

Figure 1:  Consequences of blockage after the flow cell on data.

Conclusion

Reading ‘The Art of War’, I was struck by some of Sun Tzu’s advice, military genius by all accounts. This also applies to the science of flow cytometry. In this blog I’ve tried to share with you some of this advice. We will see it applies to flow cytometry experimental design and execution. Perhaps some may see it as a stretch. But to me, anything that helps me remember and take pause as I perform my experiments is a good thing. The statements I’ve shared remind me to be alert as I go about my research efforts.  Let me leave you with one bonus quote that I think doesn’t require much explanation. “You have to believe in yourself.”

To learn more about important control measures for your flow cytometry lab, and to get access to all of our advanced materials including 20 training videos, presentations, workbooks, and private group membership, get on the Flow Cytometry Mastery Class wait list.

Join Expert Cytometry's Mastery Class

ABOUT TIM BUSHNELL, PHD

Tim Bushnell holds a PhD in Biology from the Rensselaer Polytechnic Institute. He is a co-founder of—and didactic mind behind—ExCyte, the world’s leading flow cytometry training company, which organization boasts a veritable library of in-the-lab resources on sequencing, microscopy, and related topics in the life sciences.

Tim Bushnell, PhD

Similar Articles

The Power Of Spectral Viewers And Their Use In Full Spectrum Flow Cytometry

The Power Of Spectral Viewers And Their Use In Full Spectrum Flow Cytometry

By: Tim Bushnell, PhD

What photon from yonder fluorochrome breaks?  It is … umm… hmmm. Let me see. Excitation off a 561 nm laser, with an emission maximum of 692 nm. I’m sure if Shakespeare was a flow cytometrist, he might have written that very scene. But the play is lost in time. However, since the protagonist had difficulty determining what fluorochrome was emitting photons, let’s consider how this could be figured out. In my opinion, one of the handiest flow cytometry tools is the spectral viewer. This tool helps visualize the excitation and emission profile of different fluorochromes, as well as allowing you…

Fickle Markers: Solutions For Antibody Binding Specificity Challenges

Fickle Markers: Solutions For Antibody Binding Specificity Challenges

By: Tim Bushnell, PhD

Reproducibility has been an ongoing, and important, concept in the sciences for years.  In the area of biomedical research, the alarm was sounded by several papers published in the early 2010’s.  Authors like Begley and Ellis, Prinz and coworkers, and Vasilevsky and colleagues, among others reported an alarming trend in the reproducibility of pre-clinical data.  These reports indicated between 50% to almost 90% of published pre-clinical data were not reproducible.  This was further highlighted in the article by Freedman and coworkers, who tried to identify and quantify the different sources of error that could be causing this crisis.  Figure 1,…

5 Common Flow Cytometry Questions, Answered

5 Common Flow Cytometry Questions, Answered

By: Tim Bushnell, PhD

I want to thank all of you who send us your questions about flow cytometry, so I thought I would dip into the old email bag and answer a few of the common ones here.  If your question isn’t answered this time, look for it to be answered in a future blog post.  Of course, if you want us to cover a specific topic, drop us a line.  1. How Fast Can I Go? This is  a common question. The allure of the ‘hi’ button is hard to resist.  The faster you go, the sooner you are finished with data…

Combining Flow Cytometry With Plant Science, Microorganisms, And The Environment

Combining Flow Cytometry With Plant Science, Microorganisms, And The Environment

By: Tim Bushnell, PhD

My first introduction to flow cytometry was talking to a professor who’d brought one on a research cruise to study phytoplankton. It was only later that I was introduced to the marvelous world that’s been my career for over 20 years.   In that time, I’ve had the opportunity to work with researchers in many different areas, exposing me to a wide variety of cell types and more important assays. What continues to amaze me is the number of different parameters we can measure, not just the number of fluorochromes, but the information we can extract from samples – animal, vegetable…

Common Numbers-Based Questions I Get As A Flow Cytometry Core Manager And How To Answer Them

Common Numbers-Based Questions I Get As A Flow Cytometry Core Manager And How To Answer Them

By: Tim Bushnell, PhD

Numbers are all around us.  My personal favorite is ≅1.618 aka ɸ aka ‘the golden ratio’.  It’s found throughout history, where it has influenced architects and artists. We see it in nature, in plants, and it is used in movies to frame shots. It can be approximated by the Fibonacci sequence (another math favorite of mine). However, I have not worked out how to apply this to flow cytometry.  That doesn’t mean numbers aren’t important in flow cytometry. They are central to everything we do, and in this blog, I’m going to flit around numbers-based questions that I have received…

3 Must-Have High-Dimensional Flow Cytometry Controls

3 Must-Have High-Dimensional Flow Cytometry Controls

By: Tim Bushnell, PhD

Developments such as the recent upgrade to the Cytobank analysis platform and the creation of new packages such as Immunocluster are reducing the computational expertise needed to work with high-dimensional flow cytometry datasets. Whether you are a researcher in academia, industry, or government, you may want to take advantage of the reduced barrier to entry to apply high-dimensional flow cytometry in your work. However, you’ll need the right experimental design to access the new transformative insights available through these approaches and avoid wasting the considerable time and money required for performing them. As with all experiments, a good design begins…

The Fluorochrome Less Excited: How To Build A Flow Cytometry Antibody Panel

The Fluorochrome Less Excited: How To Build A Flow Cytometry Antibody Panel

By: Tim Bushnell, PhD

Fluorochrome, antibodies and detectors are important. The journey of a thousand cells starts with a good fluorescent panel. The polychromatic panel is the combination of antibodies and fluorochromes. These will be used during the experiment to answer the biological question of interest. When you only need a few targets, the creation of the panel is relatively straightforward. It’s only when you start to get into more complex panels with multiple fluorochromes that overlap in excitation and emission gets more interesting.  FLUOROCHROMES Both full spectrum and traditional fluorescent flow cytometry rely on measuring the emission of the fluorochromes that are attached…

Flow Cytometry Year in Review: Key Changes To Know

Flow Cytometry Year in Review: Key Changes To Know

By: Meerambika Mishra

Here we are, at the end of an eventful year 2021. But with the promise of a new year 2022 to come. It has been a long year, filled with ups and downs. It is always good to reflect on the past year as we move to the future.  In Memoriam Sir Isaac Newton wrote “If I have seen further, it is by standing upon the shoulders of giants.” In the past year, we have lost some giants of our field including Zbigniew Darzynkiwicz, who contributed much in the areas of cell cycle analysis and apoptosis. Howard Shapiro, known for…

What Star Trek Taught Me About Flow Cytometry

What Star Trek Taught Me About Flow Cytometry

By: Tim Bushnell, PhD

It is no secret that I am a very big fan of the Star Trek franchise. There are many good episodes and lessons explored in the 813+ episodes, 12 movies (and counting). Don’t worry, this blog is not going to review all 813, or even 5 of them. Instead, some of the lessons I have taken away from the show that have applicability to science and flow cytometry.  “Darmok and Jalad at Tanagra.”  (ST:TNG season 5, episode 2) This is probably one of my favorite episodes, which involves Picard and an alien trying to establish a common ground and learn…

Top Industry Career eBooks

Get the Advanced Microscopy eBook

Get the Advanced Microscopy eBook

Heather Brown-Harding, PhD

Learn the best practices and advanced techniques across the diverse fields of microscopy, including instrumentation, experimental setup, image analysis, figure preparation, and more.

Get The Free Modern Flow Cytometry eBook

Get The Free Modern Flow Cytometry eBook

Tim Bushnell, PhD

Learn the best practices of flow cytometry experimentation, data analysis, figure preparation, antibody panel design, instrumentation and more.

Get The Free 4-10 Compensation eBook

Get The Free 4-10 Compensation eBook

Tim Bushnell, PhD

Advanced 4-10 Color Compensation, Learn strategies for designing advanced antibody compensation panels and how to use your compensation matrix to analyze your experimental data.