How Flow Cytometry Converts Photons To Digital Data

How exactly does a photon of light become an electron, and eventually a number in the FCS listmode file?

In flow cytometry, cells are labeled with fluorescent tags, passed by an excitation light source, and the resulting emitted photons are collected to become the data that identifies characteristics about the cells in question.

These characteristics can include what proteins are expressed, the cell cycle state, the phosphorylation state of a protein, the calcium state of the cells, levels of RNA expression, and so much more.

If there is a fluorescent probe that can measure a specific characteristic, it can often be used in flow cytometry.

A great resource for reference is the Molecular Probes Handbook, as it contains a wealth of information about different fluorescent reagents that can measure everything from apoptosis to Zn++ ion concentration.

This process involves the detection system and the electronics, and basically following a bouncing photon to its ultimate digitization.

The role of the detector on a flow cytometer is to capture photons and convert them to electric current.

By and large, the two most common detectors on commercial flow cytometers are photodiodes and photomultiplier tubes.

In a PMT, photons of light enter through a window and hit the photocathode. If the photon is of sufficient energy, they eject an electron (due to the Photoelectric Effect).

The electron is focused to one of several ‘fins’ (called dynodes), where the electrons are multiplied via secondary emission. At the end of this chain is an anode, where the electrons are converted to an electronic pulse. This is shown schematically below.

Electrons striking photoanode and creating current flow

As is shown, there is a linear relationship with the photons hitting the cathode and the output photocurrent.

It should be noted that PMTs can turn any photon that strikes the photocathode into a photocurrent.

Thus, to control the photons that are measured by any given detector, optical filters are placed in front of the PMT.

Most commonly, these are band pass filters, and as a reminder and shown below, band pass filters allow light within a given range through. To describe these filters, manufacturers label these filters with the center of the range and the size of the window of light that can pass through the filter. So a 620/40 band pass filter will allow light between 600 to 640 nm through, as shown below.

Band pass filters allow light to pass within a given range

Current produced is proportional to intensity of the fluorescent signal:

At the end of the day, one has a current produced that is proportional to the intensity of the fluorescent signal, and each PMT measures a specific wavelength range because of the optical filters placed in front of the detector.

At this point, the photons of light have been turned into electric current.

The next question is, “How does that electric current get turned into a digital value that is sorted in the FCS file?” The image below shows what a typical electronic pulse looks like.

The current produced is proportional to the intensity of the fluorescent signal

This pulse has three characteristics that can be measured: the peak height, the pulse width and the pulse area, and the integral of the height and width. The boxes represent the fact that the electronic pulse is being sampled at some frequency, and with each sampling, the values are digitized by the Analog to Digital Converter (ADC).

The ADC has two characteristics: the sampling frequency and the resolution.

The sampling frequency is expressed in megaHertz (MHZ), and ranges from 10 to 100 on the current generation of instruments. This means that the electronic pulse is being sampled between 10 to 100 million times PER SECOND.

From this sampling, the value is digitized into a range equal to 2 to the power of the ADC. This can range from 10 to 24 bits or 1,024 to 16,777,216 discrete values.

The values for some common instruments are shown below:

InstrumentVendorSampling FrequencyADC
AccuriBD Biosciences80 MHz24
FACSCanto IIBD Biosciences20 MHz14
FACSDiva   (LSR-II, Aria, Fortessa)BD Biosciences10 MHz14
FACSVerseBD Biosciences25 MHz14
GalliosBeckman-Coulter40 MHz20
MACSQuantMiltenyi Biotec??32
MoFlo XDPBeckman-Coulter100 MHz16
SH800SONY100 MHz20

It is worth noting that BD instruments use a predetermined 18 bit log-lookup table so that the range of the data is 262,144 bins.

Frequency of sampling and resolution in a Flow Cytometer:

There are two important characteristics of the ADC: the frequency of sampling and the resolution.

Natural questions might include: What are all these bits, really? Or are they bins? Or channels?

In the case of a 10 bit ADC, this means there are 1,024 discrete values that the sampled pulse can be assigned. With a higher ADC (Accuri with a 24 bit ADC), it has over 16 million discrete values.

These are the ‘channels’ or ‘bins’ and represent the actual value that has been measured from the signal pulse. So when the signal is digitized, it is assigned one of these values.

For linear data, this is pretty self-explanatory. However, with log data, this binning gets a bit tricky, especially if you consider the difference between the older FCS2 data and the current FCS3 data.

Take, for example, the FACSCalibur, which has 10 bit resolution. Since this data, when log transformed, was being measured over 4 decades, the vendor decided to spread the 1024 bins equally across the log space.

So the first decade (1-10) had 256 values, as did the second decade (10-100), the third decade (100-1000) and the fourth decade (1000-10,000). One can see that in the case of the third and fourth decades, 256 bins did not give sufficient resolution, so that each bin contained more ‘values’ than a bin in the lower decades.

On the modern digital instruments, these values are spread out to better reflect the number of values available at the higher level ADCs. Therefore, there is more resolution at higher decades than at lower decades.

Flow cytometrists tend to use bins and channels interchangeably.

Photons are captured by the detector and converted from light to electrons

Conversion of photons to electrons:

The number of bins that a given system has is based on the resolution of the ADC.

With digital instruments and the FCS 3 format, the bins are spread out to better reflect the number of possible values in the log space.

In summary, a photon is emitted from a fluorochrome. The photon is captured by the detector (e.g. the photomultiplier tube, PMT), where it is converted from light to an electron. At the same time, the PMT amplifies the incoming signal in a linear, proportional manner.

The output is an electronic signal pulse. The electronic pulses coming off the PMT is sampled many millions of times a second, and the height of that sample is digitized into a discrete value that is placed in a bin. The resolution of the ADC dictates the number of possible bins available for use.

In the end, this value is stored in the FCS file, and ready for further analysis.

To learn more about how flow cytometry converts photons to digital data, and to get access to all of our advanced materials including 20 training videos, presentations, workbooks, and private group membership, get on the Flow Cytometry Mastery Class wait list.

Join Expert Cytometry's Mastery Class

ABOUT TIM BUSHNELL, PHD

Tim Bushnell holds a PhD in Biology from the Rensselaer Polytechnic Institute. He is a co-founder of—and didactic mind behind—ExCyte, the world’s leading flow cytometry training company, which organization boasts a veritable library of in-the-lab resources on sequencing, microscopy, and related topics in the life sciences.

Tim Bushnell, PhD

Similar Articles

The Power Of Spectral Viewers And Their Use In Full Spectrum Flow Cytometry

The Power Of Spectral Viewers And Their Use In Full Spectrum Flow Cytometry

By: Tim Bushnell, PhD

What photon from yonder fluorochrome breaks?  It is … umm… hmmm. Let me see. Excitation off a 561 nm laser, with an emission maximum of 692 nm. I’m sure if Shakespeare was a flow cytometrist, he might have written that very scene. But the play is lost in time. However, since the protagonist had difficulty determining what fluorochrome was emitting photons, let’s consider how this could be figured out. In my opinion, one of the handiest flow cytometry tools is the spectral viewer. This tool helps visualize the excitation and emission profile of different fluorochromes, as well as allowing you…

Fickle Markers: Solutions For Antibody Binding Specificity Challenges

Fickle Markers: Solutions For Antibody Binding Specificity Challenges

By: Tim Bushnell, PhD

Reproducibility has been an ongoing, and important, concept in the sciences for years.  In the area of biomedical research, the alarm was sounded by several papers published in the early 2010’s.  Authors like Begley and Ellis, Prinz and coworkers, and Vasilevsky and colleagues, among others reported an alarming trend in the reproducibility of pre-clinical data.  These reports indicated between 50% to almost 90% of published pre-clinical data were not reproducible.  This was further highlighted in the article by Freedman and coworkers, who tried to identify and quantify the different sources of error that could be causing this crisis.  Figure 1,…

5 Common Flow Cytometry Questions, Answered

5 Common Flow Cytometry Questions, Answered

By: Tim Bushnell, PhD

I want to thank all of you who send us your questions about flow cytometry, so I thought I would dip into the old email bag and answer a few of the common ones here.  If your question isn’t answered this time, look for it to be answered in a future blog post.  Of course, if you want us to cover a specific topic, drop us a line.  1. How Fast Can I Go? This is  a common question. The allure of the ‘hi’ button is hard to resist.  The faster you go, the sooner you are finished with data…

Combining Flow Cytometry With Plant Science, Microorganisms, And The Environment

Combining Flow Cytometry With Plant Science, Microorganisms, And The Environment

By: Tim Bushnell, PhD

My first introduction to flow cytometry was talking to a professor who’d brought one on a research cruise to study phytoplankton. It was only later that I was introduced to the marvelous world that’s been my career for over 20 years.   In that time, I’ve had the opportunity to work with researchers in many different areas, exposing me to a wide variety of cell types and more important assays. What continues to amaze me is the number of different parameters we can measure, not just the number of fluorochromes, but the information we can extract from samples – animal, vegetable…

Common Numbers-Based Questions I Get As A Flow Cytometry Core Manager And How To Answer Them

Common Numbers-Based Questions I Get As A Flow Cytometry Core Manager And How To Answer Them

By: Tim Bushnell, PhD

Numbers are all around us.  My personal favorite is ≅1.618 aka ɸ aka ‘the golden ratio’.  It’s found throughout history, where it has influenced architects and artists. We see it in nature, in plants, and it is used in movies to frame shots. It can be approximated by the Fibonacci sequence (another math favorite of mine). However, I have not worked out how to apply this to flow cytometry.  That doesn’t mean numbers aren’t important in flow cytometry. They are central to everything we do, and in this blog, I’m going to flit around numbers-based questions that I have received…

3 Must-Have High-Dimensional Flow Cytometry Controls

3 Must-Have High-Dimensional Flow Cytometry Controls

By: Tim Bushnell, PhD

Developments such as the recent upgrade to the Cytobank analysis platform and the creation of new packages such as Immunocluster are reducing the computational expertise needed to work with high-dimensional flow cytometry datasets. Whether you are a researcher in academia, industry, or government, you may want to take advantage of the reduced barrier to entry to apply high-dimensional flow cytometry in your work. However, you’ll need the right experimental design to access the new transformative insights available through these approaches and avoid wasting the considerable time and money required for performing them. As with all experiments, a good design begins…

The Fluorochrome Less Excited: How To Build A Flow Cytometry Antibody Panel

The Fluorochrome Less Excited: How To Build A Flow Cytometry Antibody Panel

By: Tim Bushnell, PhD

Fluorochrome, antibodies and detectors are important. The journey of a thousand cells starts with a good fluorescent panel. The polychromatic panel is the combination of antibodies and fluorochromes. These will be used during the experiment to answer the biological question of interest. When you only need a few targets, the creation of the panel is relatively straightforward. It’s only when you start to get into more complex panels with multiple fluorochromes that overlap in excitation and emission gets more interesting.  FLUOROCHROMES Both full spectrum and traditional fluorescent flow cytometry rely on measuring the emission of the fluorochromes that are attached…

Flow Cytometry Year in Review: Key Changes To Know

Flow Cytometry Year in Review: Key Changes To Know

By: Meerambika Mishra

Here we are, at the end of an eventful year 2021. But with the promise of a new year 2022 to come. It has been a long year, filled with ups and downs. It is always good to reflect on the past year as we move to the future.  In Memoriam Sir Isaac Newton wrote “If I have seen further, it is by standing upon the shoulders of giants.” In the past year, we have lost some giants of our field including Zbigniew Darzynkiwicz, who contributed much in the areas of cell cycle analysis and apoptosis. Howard Shapiro, known for…

What Star Trek Taught Me About Flow Cytometry

What Star Trek Taught Me About Flow Cytometry

By: Tim Bushnell, PhD

It is no secret that I am a very big fan of the Star Trek franchise. There are many good episodes and lessons explored in the 813+ episodes, 12 movies (and counting). Don’t worry, this blog is not going to review all 813, or even 5 of them. Instead, some of the lessons I have taken away from the show that have applicability to science and flow cytometry.  “Darmok and Jalad at Tanagra.”  (ST:TNG season 5, episode 2) This is probably one of my favorite episodes, which involves Picard and an alien trying to establish a common ground and learn…

Top Industry Career eBooks

Get the Advanced Microscopy eBook

Get the Advanced Microscopy eBook

Heather Brown-Harding, PhD

Learn the best practices and advanced techniques across the diverse fields of microscopy, including instrumentation, experimental setup, image analysis, figure preparation, and more.

Get The Free Modern Flow Cytometry eBook

Get The Free Modern Flow Cytometry eBook

Tim Bushnell, PhD

Learn the best practices of flow cytometry experimentation, data analysis, figure preparation, antibody panel design, instrumentation and more.

Get The Free 4-10 Compensation eBook

Get The Free 4-10 Compensation eBook

Tim Bushnell, PhD

Advanced 4-10 Color Compensation, Learn strategies for designing advanced antibody compensation panels and how to use your compensation matrix to analyze your experimental data.