What Is The Coulter Principle And Why You Need To Obtain Complete Blood Counts By Flow Cytometry

When you get your blood drawn for your annual physical, or if you have been hospitalized, there are some common flow cytometry tests that your Medical Technician will run.

These tests will often analyze your blood lipids, vitamin levels, ion concentrations, and more often than not, your complete blood count (CBC).

The CBC is an automated hematology test (really several tests being run in the same instrument) that looks at the levels of all the cells in your blood, providing your physician with valuable information about your health.

The CBC is a powerful addition to many flow cytometry workflows.

Using just a small sample of blood, the CBC generates an extensive amount of information WITHOUT the need for centrifugation or multi-color staining experiments. Running a CBC is fast, easy, and inexpensive.

What Is The Coulter Principle?

The CBC generates accurate cell counts using the Coulter principle.

Named after its discoverer, Wallace Coulter, the Coulter principle states that particles passing through an orifice (along with an electrical current) will produce an increase in impedance, due to the displacement of electrolytes caused by the presence of the particle.

This change in impedance is proportional to the volume of the particle. The Coulter principle has been used for particle counting and sizing in a variety of fields, but one of the first and most impacted fields was hematology.

Labs that once took an hour to produce a blood count from a single patient using a microscope were now produced an order of magnitude faster and more consistently.

What Does A Complete Blood Count (CBC) Tell You?

A CBC, while inexpensive to run, gives you very valuable information. In addition, some CBC instruments only require a few microliters of blood to obtain an accurate reading.

For example, the Abbott Cell-Dyn 3500 requires about 200 µl of blood, but some newer instruments only require 15-20 µl of blood. 

A typical CBC report will provide you with the following information…

1. Leukocyte count with differential—total white blood cell count and a 5 part differential (lymphocytes, monocytes, neutrophils, basophils, and eosinophils). Reported as total counts per microliter of blood and as a percentage of total leukocytes for the differential.

Depending on the instrument, this might only be a three part differential (lymphs, monos, and granulocytes). Instruments with optical counting in addition to impedance counting can give the 5 part differential.

2. Red blood cell count (also known as corpuscles).

3. Platelet count (also known as thrombocytes)—critical for blood clotting.

4. Platelet volume (measured in femtoliters).

5. Hemoglobin concentration (measured in grams per deciliter).

6. Hematocrit (also known as packed cell volume, expressed as a percentage measuring the total amount).

7. How much blood is cell volume versus plasma volume (expressed as a percentage).

8. Mean corpuscular volume (average volume of red blood cells).

9. Mean corpuscular hemoglobin and mean corpuscular hemoglobin concentration—these values are arrived at arithmetically using the red cell volume and hemoglobin results.

10. Red cell distribution width—variation in volume of RBCs.

complete blood count numbers | Expert Cytometry | coulter principle

How Can A CBC Help Me With My Flow Cytometry Experiments?

From your physician’s perspective, the CBC can reveal infections that might not yet be causing symptoms.

For example, elevated neutrophils could indicate a bacterial infection, elevated B and T lymphocyte function could indicate some viral infections or leukemia, elevated eosinophils could indicate a parasitic infection or an allergic reaction, and low red blood cells could indicate iron deficient anemia.

In the world of clinical research, a CBC should always be run on the human clinical research samples.

As a result, any obvious outliers can be removed from the study, reducing the spread of the data and reducing the risk of confounding your interpretation of the data. For investigators that wish to express their data in terms of count per volume of blood, running a CBC is one of the best ways to obtain accurate values.

You can lose somewhere between 10% and 20% of your cells with each centrifugation step, and that amount will vary depending on your buffer, cells, and whether or not you’re using glass or plastic tubes. Using counting beads is great for generating a total sample cell count, however, if your goal is to ascertain a whole blood count, it’s best to apply your gated fluorescent data statistics to the white blood cell count per microliter of total blood.

More Benefits Of Obtaining A CBC By Flow Cytometry

In addition to the enhanced accuracy and confidence a CBC provides, it’s also very fast.

Counting lysed blood via hemacytometer is tedious at best and performing it for more than a handful of samples is excessively time-consuming. Even image-based hand-held impedance counters are time-consuming, not to mention their decreased accuracy due to counting after lysis and centrifugation.) A Cell-Dyn or similar instrument, on the other hand, provides accurate results in 30 seconds.

Your mouse experiments can also benefit from CBCs. Mice with active infections might not be identified by animal tech welfare examinations but their increased neutrophil activity would certainly be identified by a flow cytometric CBC.

Another nice feature of many CBC instruments is the ability to set reference ranges for each reported value. In this way, low and high values are “flagged” on screen and in printed reports, allowing you to rapidly visualize and investigate any suspicious outliers.

CBCs also provide an additional layer of quality control in terms of how many cells you’re losing during your subsequent staining protocols. If you noticed that you’re losing more cells than you should be, you can investigate the problem by ensuring that everyone is following their standard operating procedures for lysis and centrifugation, testing a new lot of antibodies and other reagents, and so on.

Assuming you can spare the blood volume required, the CBC is a fast, easy, and inexpensive way to get additional data for your experiments. CBC experiments will allow you to quickly identify infections, outliers, and a myriad of cellular information. If you haven’t already invested in a CBC instrument, they are commonly sold on the refurbished market. Finally, in terms of cost per test, CBC reagents are also inexpensive, while quality control reagents are very reasonable as well.

To learn more about current best practices in flow cytometry, and to get access to all of our advanced materials including 20 training videos, presentations, workbooks, and private group membership, get on the Flow Cytometry Mastery Class wait list.

Join Expert Cytometry's Mastery Class

ABOUT TIM BUSHNELL, PHD

Tim Bushnell holds a PhD in Biology from the Rensselaer Polytechnic Institute. He is a co-founder of—and didactic mind behind—ExCyte, the world’s leading flow cytometry training company, which organization boasts a veritable library of in-the-lab resources on sequencing, microscopy, and related topics in the life sciences.

Tim Bushnell, PhD

Similar Articles

The Power Of Spectral Viewers And Their Use In Full Spectrum Flow Cytometry

The Power Of Spectral Viewers And Their Use In Full Spectrum Flow Cytometry

By: Tim Bushnell, PhD

What photon from yonder fluorochrome breaks?  It is … umm… hmmm. Let me see. Excitation off a 561 nm laser, with an emission maximum of 692 nm. I’m sure if Shakespeare was a flow cytometrist, he might have written that very scene. But the play is lost in time. However, since the protagonist had difficulty determining what fluorochrome was emitting photons, let’s consider how this could be figured out. In my opinion, one of the handiest flow cytometry tools is the spectral viewer. This tool helps visualize the excitation and emission profile of different fluorochromes, as well as allowing you…

3 Must-Have High-Dimensional Flow Cytometry Controls

3 Must-Have High-Dimensional Flow Cytometry Controls

By: Tim Bushnell, PhD

Developments such as the recent upgrade to the Cytobank analysis platform and the creation of new packages such as Immunocluster are reducing the computational expertise needed to work with high-dimensional flow cytometry datasets. Whether you are a researcher in academia, industry, or government, you may want to take advantage of the reduced barrier to entry to apply high-dimensional flow cytometry in your work. However, you’ll need the right experimental design to access the new transformative insights available through these approaches and avoid wasting the considerable time and money required for performing them. As with all experiments, a good design begins…

The Fluorochrome Less Excited: How To Build A Flow Cytometry Antibody Panel

The Fluorochrome Less Excited: How To Build A Flow Cytometry Antibody Panel

By: Tim Bushnell, PhD

Fluorochrome, antibodies and detectors are important. The journey of a thousand cells starts with a good fluorescent panel. The polychromatic panel is the combination of antibodies and fluorochromes. These will be used during the experiment to answer the biological question of interest. When you only need a few targets, the creation of the panel is relatively straightforward. It’s only when you start to get into more complex panels with multiple fluorochromes that overlap in excitation and emission gets more interesting.  FLUOROCHROMES Both full spectrum and traditional fluorescent flow cytometry rely on measuring the emission of the fluorochromes that are attached…

Flow Cytometry Year in Review: Key Changes To Know

Flow Cytometry Year in Review: Key Changes To Know

By: Meerambika Mishra

Here we are, at the end of an eventful year 2021. But with the promise of a new year 2022 to come. It has been a long year, filled with ups and downs. It is always good to reflect on the past year as we move to the future.  In Memoriam Sir Isaac Newton wrote “If I have seen further, it is by standing upon the shoulders of giants.” In the past year, we have lost some giants of our field including Zbigniew Darzynkiwicz, who contributed much in the areas of cell cycle analysis and apoptosis. Howard Shapiro, known for…

What Star Trek Taught Me About Flow Cytometry

What Star Trek Taught Me About Flow Cytometry

By: Tim Bushnell, PhD

It is no secret that I am a very big fan of the Star Trek franchise. There are many good episodes and lessons explored in the 813+ episodes, 12 movies (and counting). Don’t worry, this blog is not going to review all 813, or even 5 of them. Instead, some of the lessons I have taken away from the show that have applicability to science and flow cytometry.  “Darmok and Jalad at Tanagra.”  (ST:TNG season 5, episode 2) This is probably one of my favorite episodes, which involves Picard and an alien trying to establish a common ground and learn…

5 Flow Cytometry Strategies That Sun Tzu Taught Me

5 Flow Cytometry Strategies That Sun Tzu Taught Me

By: Tim Bushnell, PhD

Sun Tzu was a Chinese general and philosopher. His most famous writing is ‘The Art of War’, and has been studied by generals and CEOs, to glean ideas and strategies to help their missions. I was recently rereading this work and thought to myself if any of Sun Tzu’s lessons could apply to flow cytometry.  So I have identified 5 points that I think lend themselves to thinking about flow cytometry.  “Quickness is the essence of the war.” In flow cytometry, speed is of the essence. The longer the cells are out of their natural environment, the less happy they…

A Basic Guide To Flow Cytometry (3 Foundational Concepts)

A Basic Guide To Flow Cytometry (3 Foundational Concepts)

By: Meerambika Mishra

Mastering foundational concepts are imperative for successfully using any technique or system.  Robert Heinlein introduced the term ‘Grok’  in his novel Stranger in a Strange Land. Ever since then it has made its way into popular culture. To Grok something is to understand it intuitively, fully. As a cytometrist, there are several key concepts that you must grok to be successful in your career. These foundational concepts are the key tools that we use day in and day out to identify and characterize our cells of interest.  Cells Flow cytometry measures biological processes at the whole cell level. To do…

Which Fluorophores To Use For Your Microscopy Experiment

Which Fluorophores To Use For Your Microscopy Experiment

By: Heather Brown-Harding, PhD

Fluorophore selection is important. I have often been asked by my facility users which fluorophore is best suited for their experiments. The answer to this is mostly dependent on whether they are using a widefield microscope with set excitation/emission cubes or a laser based system that lets you select the laser and the emission window. Once you have narrowed down which fluorophores you can excite and collect the correct emission, you can further refine the specific fluorophore that is best for your experiment.  In this blog  we will discuss how to determine what can work with your microscope, and how…

4 No Cost Ways To Improve Your Microscopy Image Quality

4 No Cost Ways To Improve Your Microscopy Image Quality

By: Heather Brown-Harding, PhD

Image quality is critical for accurate and reproducible data. Many people get stuck on the magnification of the objective or on using a confocal instead of a widefield microscope. There are several other factors that affect the image quality such as the numerical aperture of the objective, the signal-to-noise ratio of the system, or the brightness of the sample.  Numerical aperture is the ability of an objective to collect light from a sample, but it contributes to two key formulas that will affect your image quality. The first is the theoretical resolution of the objective. It is expressed with the…

Top Industry Career eBooks

Get the Advanced Microscopy eBook

Get the Advanced Microscopy eBook

Heather Brown-Harding, PhD

Learn the best practices and advanced techniques across the diverse fields of microscopy, including instrumentation, experimental setup, image analysis, figure preparation, and more.

Get The Free Modern Flow Cytometry eBook

Get The Free Modern Flow Cytometry eBook

Tim Bushnell, PhD

Learn the best practices of flow cytometry experimentation, data analysis, figure preparation, antibody panel design, instrumentation and more.

Get The Free 4-10 Compensation eBook

Get The Free 4-10 Compensation eBook

Tim Bushnell, PhD

Advanced 4-10 Color Compensation, Learn strategies for designing advanced antibody compensation panels and how to use your compensation matrix to analyze your experimental data.