7 Advanced Flow Cytometry Data Analysis Tips For Multi-Color Experiments

There was a time when two- and three-color experiments were considered very complex flow cytometry experiments.

A limited number of dyes and light sources meant a limited number of options overall. 

In fact, the first experiments sorting cells with fluorescently-labeled antibodies were performed by Len Herzenberg in 1972 and could only detect fluorescence from an Argon laser source above 530 nm. These early experiments were performed using rhodamine- and fluorescein-tagged antibodies.

Times have changed. Now, we have spatially separated lasers and a seemingly unlimited number of dyes.

From 2-Colors To 10-Colors

Not long after Herzenberg’s work, Howard Shapiro began designing a series of multiparameter instruments. While instrumentation was evolving, there was parallel development in fluorescent colors to allow for detection of more antigens.

For example, the patent for the phycobiliprotein uses in FACS and other applications was issued to Glazer, Oi, and Stryer in 1995. Then, in the early 2000’s, Alexa Fluor dyes came onto the scene.

In today’s world, many scientists have access to instruments capable of running multi-color experiments with 10 or more colors. 

The leap from 2 to 10 may seem small, but there are many factors to consider in the design and analysis of experiments that make full use of instruments that can handle these additional colors.

Each additional fluorochrome (potentially) affects each other fluorochrome, which makes your analysis more and more complex.

7 10-Color Flow Cytometry Tips

Imagine analyzing a 2-color experiment. With 2 biaxial plots and a single quadrant gate, you have only 4 populations to report.

Now add a 3rd color. By doing so, you’ve increased your population count to 8. With 4-colors, you’ve increased your population count to 16.

On and on it goes until you get to 10-colors. Now you have 1024 possible combinations!

With this kind of complexity, careful experimental planning is not a luxury, it’s a necessity. Here are 7 tips for preparing and analyzing 10-color flow cytometry experiments…

1. Understand compensation and data spread.

To build a 10-color panel, you’re going to have to put more than one color on a single laser line and its detectors. This means you will have to perform compensation. There’s no way to avoid it.

The key point to emphasize for these complex experiments is that compensation reveals spread in your data. The spread can make it more difficult to separate dim positive signals from negative signals.

This can be observed in very simple examples, such as the one below where antibody capture beads have been properly compensated with an APC-conjugated antibody. Using a straight quadrant gate based on the unstained bead population, the user mistakenly identifies 10% of those beads as APC-Cy7 positive.

The true shape of the data is better matched with a curved quadrant that takes into account the spread in the APC-Cy7 channel.

2. Remember to consider fluorochrome brightness and antigen density. 

For most experiments, it’s important to pair your lowest expressing targets with your brightest fluorochromes. This is all the more important when doing 10-color experiments. 

As mentioned in point #1 above, compensation creates spread in your data. If you make the mistake of placing a lowly expressing marker on a channel with many overlapping signals, it will be very difficult to resolve the resulting dimly positive signal. Even highly expressed antigens become more difficult to discern the more colors you add to your experiment.

3. Use FMO controls to help deal with spectral overlap. 

The best way to deal with the spread associated with spectral overlap in multi-color experiments is to make liberal use of Fluorescence Minus One, or FMO controls. Not only do FMO controls save you from drawing incorrect gates, they can save you from poorly compensated data. 

In the example below (courtesy of Bridget McLaughlin, UC Davis Core Facility), pay particular attention to the plot showing the activation of CD8+ cells. As the note in this area of the plot indicates, it can be very difficult to plot’s quadrants, as they do not separate as clearly as the plots of the mature cells. FMOs are critical for removing ambiguity and human error during the data analysis portion of your experiment.

4. Use biological controls where appropriate.

If your 10-color panel includes cell signaling or other intracellular staining, make sure you include an unstimulated control along with your FMO control. Very often, there will be baseline expression levels of cytokines and/or phosphoproteins in your cells. The FMO will allow you to determine this baseline level of activity, while the biological control allows you to see the effect of your treatment.

5. Gate out populations before gating on your population of interest.

Once you begin analyzing your data, try not to go get locked into looking at your data in one way. The fact that you know exactly what you’re looking for can limit your perspective.

For example, if you’re interested in a B cell population, don’t just gate on your B cells, analyze them, and call it a day. Instead, try gating out the populations you’re not interested in first.

Returning to our example, if you’re interested in B cells, gate out T cells first, then look at the population again. Next, gate out monocytes, then look at your population again. You might be surprised by what you discover.

Cells of different shapes and sizes can have different levels of autofluorescence so removing them from further analysis can allow you to be more aggressive in gating for dim levels of other markers.

6. Trying multiple combinations of dyes to get the best results.

For most 10-color experiments, you should try several combinations of fluorochromes to get the best possible separation of signals.

You can make several inferences about which marker should be placed on which fluorochrome using what you know about your experiment, sample type, and instrument configurations. Still, during the experiment itself, thing may not come out as predicted by your design.

As seen in the example below (courtesy of Mario Roederer, National Institutes of Health), it can be very helpful to try out a few combinations of staining cocktails during the initial set-up of your experiment to ensure the populations of interest will be well-defined. Depending on what fluorochrome is used for these 2 markers you can completely lose resolution of the 4 populations, or make it very easy to see.

7. Use OMIPs as a starting point for your experiment. 

By looking in the OMIPs section in Cytometry A, you will find validated protocols you can use as a backbone for your own experiments.

One thing you will notice is that OMIPs are short and direct. This is what makes them such a great resource—you don’t have to dig around in the text and methods sections of papers to see what was done.

OMIPs provide figures with examples of all gates already drawn and the purpose of each gate, a summary describing the purpose of the panel and cross-referencing each panel to similar OMIPs, a short background discussion of why certain markers were used or not used, and finally, a reagent list.

The trend of the last 4 decades of flow cytometric instrumentation has been toward more colors for less money. However, this has also made it easier to generate ugly-looking data that is difficult to interpret and even more difficult to publish. When your own multi-color experiments start calling for more and more antigens in your design apply these tips and waste no time, money, and effort on bad data.

To learn more about performing 10+ color experiments and to get access to all of our advanced materials including 20 training videos, presentations, workbooks, and private group membership, get on the Flow Cytometry Mastery Class wait list.

Join Expert Cytometry's Mastery Class

ABOUT TIM BUSHNELL, PHD

Tim Bushnell holds a PhD in Biology from the Rensselaer Polytechnic Institute. He is a co-founder of—and didactic mind behind—ExCyte, the world’s leading flow cytometry training company, which organization boasts a veritable library of in-the-lab resources on sequencing, microscopy, and related topics in the life sciences.

Tim Bushnell, PhD

Similar Articles

Fickle Markers: Solutions For Antibody Binding Specificity Challenges

Fickle Markers: Solutions For Antibody Binding Specificity Challenges

By: Tim Bushnell, PhD

Reproducibility has been an ongoing, and important, concept in the sciences for years.  In the area of biomedical research, the alarm was sounded by several papers published in the early 2010’s.  Authors like Begley and Ellis, Prinz and coworkers, and Vasilevsky and colleagues, among others reported an alarming trend in the reproducibility of pre-clinical data.  These reports indicated between 50% to almost 90% of published pre-clinical data were not reproducible.  This was further highlighted in the article by Freedman and coworkers, who tried to identify and quantify the different sources of error that could be causing this crisis.  Figure 1,…

The Fluorochrome Less Excited: How To Build A Flow Cytometry Antibody Panel

The Fluorochrome Less Excited: How To Build A Flow Cytometry Antibody Panel

By: Tim Bushnell, PhD

Fluorochrome, antibodies and detectors are important. The journey of a thousand cells starts with a good fluorescent panel. The polychromatic panel is the combination of antibodies and fluorochromes. These will be used during the experiment to answer the biological question of interest. When you only need a few targets, the creation of the panel is relatively straightforward. It’s only when you start to get into more complex panels with multiple fluorochromes that overlap in excitation and emission gets more interesting.  FLUOROCHROMES Both full spectrum and traditional fluorescent flow cytometry rely on measuring the emission of the fluorochromes that are attached…

Flow Cytometry Year in Review: Key Changes To Know

Flow Cytometry Year in Review: Key Changes To Know

By: Meerambika Mishra

Here we are, at the end of an eventful year 2021. But with the promise of a new year 2022 to come. It has been a long year, filled with ups and downs. It is always good to reflect on the past year as we move to the future.  In Memoriam Sir Isaac Newton wrote “If I have seen further, it is by standing upon the shoulders of giants.” In the past year, we have lost some giants of our field including Zbigniew Darzynkiwicz, who contributed much in the areas of cell cycle analysis and apoptosis. Howard Shapiro, known for…

What Star Trek Taught Me About Flow Cytometry

What Star Trek Taught Me About Flow Cytometry

By: Tim Bushnell, PhD

It is no secret that I am a very big fan of the Star Trek franchise. There are many good episodes and lessons explored in the 813+ episodes, 12 movies (and counting). Don’t worry, this blog is not going to review all 813, or even 5 of them. Instead, some of the lessons I have taken away from the show that have applicability to science and flow cytometry.  “Darmok and Jalad at Tanagra.”  (ST:TNG season 5, episode 2) This is probably one of my favorite episodes, which involves Picard and an alien trying to establish a common ground and learn…

5 Flow Cytometry Strategies That Sun Tzu Taught Me

5 Flow Cytometry Strategies That Sun Tzu Taught Me

By: Tim Bushnell, PhD

Sun Tzu was a Chinese general and philosopher. His most famous writing is ‘The Art of War’, and has been studied by generals and CEOs, to glean ideas and strategies to help their missions. I was recently rereading this work and thought to myself if any of Sun Tzu’s lessons could apply to flow cytometry.  So I have identified 5 points that I think lend themselves to thinking about flow cytometry.  “Quickness is the essence of the war.” In flow cytometry, speed is of the essence. The longer the cells are out of their natural environment, the less happy they…

A Basic Guide To Flow Cytometry (3 Foundational Concepts)

A Basic Guide To Flow Cytometry (3 Foundational Concepts)

By: Meerambika Mishra

Mastering foundational concepts are imperative for successfully using any technique or system.  Robert Heinlein introduced the term ‘Grok’  in his novel Stranger in a Strange Land. Ever since then it has made its way into popular culture. To Grok something is to understand it intuitively, fully. As a cytometrist, there are several key concepts that you must grok to be successful in your career. These foundational concepts are the key tools that we use day in and day out to identify and characterize our cells of interest.  Cells Flow cytometry measures biological processes at the whole cell level. To do…

Which Fluorophores To Use For Your Microscopy Experiment

Which Fluorophores To Use For Your Microscopy Experiment

By: Heather Brown-Harding, PhD

Fluorophore selection is important. I have often been asked by my facility users which fluorophore is best suited for their experiments. The answer to this is mostly dependent on whether they are using a widefield microscope with set excitation/emission cubes or a laser based system that lets you select the laser and the emission window. Once you have narrowed down which fluorophores you can excite and collect the correct emission, you can further refine the specific fluorophore that is best for your experiment.  In this blog  we will discuss how to determine what can work with your microscope, and how…

4 No Cost Ways To Improve Your Microscopy Image Quality

4 No Cost Ways To Improve Your Microscopy Image Quality

By: Heather Brown-Harding, PhD

Image quality is critical for accurate and reproducible data. Many people get stuck on the magnification of the objective or on using a confocal instead of a widefield microscope. There are several other factors that affect the image quality such as the numerical aperture of the objective, the signal-to-noise ratio of the system, or the brightness of the sample.  Numerical aperture is the ability of an objective to collect light from a sample, but it contributes to two key formulas that will affect your image quality. The first is the theoretical resolution of the objective. It is expressed with the…

What Is Total Internal Reflection Fluorescence (TIRF) Microscopy & Is It Right For You?

What Is Total Internal Reflection Fluorescence (TIRF) Microscopy & Is It Right For You?

By: Heather Brown-Harding, PhD

TIRF is not as common as other microscopy based techniques due to certain restrictions. We will discuss these restrictions, then analyze why it might be perfect for your experiment.  TIRF relies on an evanescent wave, created through a critical angle of coherent light (i.e. laser) that reaches a refractive index mismatch.  What does it mean in practice?  A high angle laser reflects off the interface of the coverslip and the sample. Although the depth that this wave penetrates is dependent on the wavelength of the light, in practice it is approximately 50-300nm from the coverslip. Therefore, the cell membrane is…

Top Industry Career eBooks

Get the Advanced Microscopy eBook

Get the Advanced Microscopy eBook

Heather Brown-Harding, PhD

Learn the best practices and advanced techniques across the diverse fields of microscopy, including instrumentation, experimental setup, image analysis, figure preparation, and more.

Get The Free Modern Flow Cytometry eBook

Get The Free Modern Flow Cytometry eBook

Tim Bushnell, PhD

Learn the best practices of flow cytometry experimentation, data analysis, figure preparation, antibody panel design, instrumentation and more.

Get The Free 4-10 Compensation eBook

Get The Free 4-10 Compensation eBook

Tim Bushnell, PhD

Advanced 4-10 Color Compensation, Learn strategies for designing advanced antibody compensation panels and how to use your compensation matrix to analyze your experimental data.