4 Critical Components In Cellular Proliferation Measurement

Cellular proliferation is a critical component in biological systems.

While normal cell proliferation keeps the body functioning, abnormal proliferation (such as in cancer) can be a target for therapy.  

Measuring how cells proliferate in response to a stimulus is a time-honored assay in science. This can be as simple as a cell count between untreated and treated cells. More sophisticated assays can include the use of 3H-thymidine or colorimetric assays (MTT assays).

While these all can measure proliferation, they lack the finesse that flow cytometry can bring to the assay – which allows the phenotypic identification of which cells are actually dividing, as well as allowing for calculations of values such as the precursor frequency, the percentage of cells that have divided, a proliferation index, and more.

Proliferation calls for cells to make a trip around the cell cycle, and there are many ways to measure cell division.

The focus here is on the long-term measure of cell division—the ‘temporal’ dimension for measuring such biological processes as:

  • Proliferation of immune cells in response to stimulation
  • Self-renewal of stem cells
  • Biological homeostasis
  • Tumor cell proliferation

To this end, there are several critical components in developing, validating and optimizing an assay to make these measures using flow cytometry. These 4 components are…

1. Pick the right cellular proliferation dye.

Determine which dye you want to use for proliferation. The qualities of a good dye for proliferation include:

  • It is taken up by live cells
  • It stains brightly
  • It is well retained by the cells
  • It segregates equally between daughter cells

There are two major classes of these dyes, based on where the dye is retained by the cells.  

The first class is the intracellular dyes that enter the cell, are acted upon by cellular esterases which cleave the compound into a fluorescent form that can also interact with intracellular molecules, thus binding inside the cell.

The most common of these dyes is carboxyfluorescein diacetate succinimidyl ester (or CFDA-SE).  When this enters the cell, it is cleaved to the active form CFSE, which is amine reactive, binding to intracellular proteins, and has been used extensively for cell tracking and proliferation.

The second class of dyes for proliferation are lipophilic dyes that bind to the cell membrane.  

These dyes are typically not fluorescent until they are incorporated into the cellular membrane, and over time distribute over the whole cell. There are a host of these dyes, one of the most popular is PKH26. Table 1 below lists some of the most common cell proliferation dyes available.

There are many more dyes available, and a quick search of your favorite vendor’s catalog will reveal one that will work for your needs. The Molecular Probes Handbook is especially useful in selecting a dye for your experiment needs and instrument capabilities.

Table 1:  Some Common Cell Proliferation Dyes

cellular proliferation measurement | Expert Cytometry | measuring cellular proliferation

2.  Validate your cellular proliferation dye—make sure your cells like it!

After you have selected your dye, it is critical to optimize the labeling reaction for the dye.

This will include optimizing the labeling solution (typically between 0.1 to 10 μM), the cell concentration (between 1-20 million cells per ml), the incubation time and temperature, and the quenching step.

In the case of succinimidyl dyes, this would involve adding protein (BSA) and letting the cells rest for 5-10 minutes before washing. As part of this assay development, it is important to make sure that the dye does not kill the cells.

After labeling, a viability check is critical.  

Shown below are data from Dr. Andy Filby (head of flow cytometry at Newcastle University): cells were labeled with increasing amounts of CellTrace Violet (CTV), CFSE and eFluor670 (EPD) and the viability measure.  At 1 μM, the cells have about the same viability, but this rapidly changes, especially for CFSE, where increasing amounts of the dye increased cell death.

cellular proliferation measurement | Expert Cytometry | measuring cellular proliferation

More dye is better, in so much as the brighter the signal, the more generations can be measured in the experiment, but not at the expense of increased dead cells.

Dead cells tell no tales.

3.  Optimize your flow cytometry instrument.

In an ideal world, as the cells divide, the fluorescence signal would decrease precisely by ½ and calculating the proliferation metrics would be easy.

As shown in this figure, modified from a lecture Dr. Andy Filby presented for Expert Cytometry (and available to Mastery Class members here), the reality is not that pretty.

cellular proliferation measurement | Expert Cytometry | measuring cellular proliferation

Several factors influence the spread of the data, including:

  • The true biological variance inherent in the cell systems
  • The intrinsic spread because of fluorochrome labeling and speed of the cells through the system
  • The extrinsic spread because of the instrument (such as optics, laser power, alignment)

The consequences of these factors can be reduced in several ways:

  • Careful planning of the experiments
  • Run the cells at low differential pressure
  • Monitor the laser alignment
  • Keep the instrument (flow cell) clean

Before running samples, run a standard bead set to validate the instrument. Checking linearity, sensitivity, and alignment before the actual samples are run is a good way to help minimize the causes that can be controlled.

4.  Analyze your proliferation data correctly.

In designing a proliferation experiment, as with any flow cytometry experiment, it’s important to develop a data analysis plan before beginning the experiments.

There are three informative parameters that can be measured from a properly constructed experiment. These values are:

  • The Percent Divided – the measure of the percent of the input cells that entered division.
  • The Division Index – a measure of the average number of divisions which includes the undivided cells.
  • The Proliferation Index – the average number of divisions that exclude the undivided cells.

There is a clear temptation to manually gate on the populations and calculate these values, because of the overlap between peaks. It is impossible to easily set the gates to avoid the overlap. This is where modeling of the data comes into play.

Take, for example, the below data (courtesy of Dr. Andy Filby)…

On the left, the populations have been manually gated, while on the right, they have been fitted to a model using the FlowJo™ proliferation package.

Manually fitting the data gives a percent divided at 78%, while the model fitted data shows a percent divided of only 59%.

cellular proliferation measurement | Expert Cytometry | measuring cellular proliferation

It’s clear that the above difference has a major consequence on the interpretation of the data.  

It’s also important to remember dyes like CFSE have a 24-48 hour proliferation-independent loss of signal that must be taken into account before measuring proliferation.

Another thing to remember is that as the cells divide, it will become over-represented in the data by 2division round.  To determine the true percentage of dividing cells, you have to do some math. This starts by correcting the frequency of each generation by dividing by 2division round.

These values are then added together, and the data normalized by that value to determine the real frequency in each population. Furthermore, by subtracting the undivided fraction from 100, it is possible to get the precursor frequency.

Making sure that you know the data needed to answer the biological questions, and the power (and limitations) of the analysis of the data is critical.

Define the question, define the statistics needed and then, and only then, foray into the lab and begin the experiments.

Proliferation assays are a powerful tool for understanding and monitoring this important cell process. Understanding this process, how it can be dysregulated and what ways it can be controlled (chemically), is a critical process. The development of new treatments for cancer, for example, target the proliferation of the cancer cells. Failure to properly design and analyze the data will result in missed opportunities and false leads. Knowing the steps to optimize these assays and properly interpret the results, as discussed here, will help ensure the best data and best opportunities are pursued.

To learn more about getting your flow cytometry data published and to get access to all of our advanced materials including 20 training videos, presentations, workbooks, and private group membership, get on the Flow Cytometry Mastery Class wait list.

Join Expert Cytometry's Mastery Class

ABOUT TIM BUSHNELL, PHD

Tim Bushnell holds a PhD in Biology from the Rensselaer Polytechnic Institute. He is a co-founder of—and didactic mind behind—ExCyte, the world’s leading flow cytometry training company, which organization boasts a veritable library of in-the-lab resources on sequencing, microscopy, and related topics in the life sciences.

Tim Bushnell, PhD

Similar Articles

Common Numbers-Based Questions I Get As A Flow Cytometry Core Manager And How To Answer Them

Common Numbers-Based Questions I Get As A Flow Cytometry Core Manager And How To Answer Them

By: Tim Bushnell, PhD

Numbers are all around us.  My personal favorite is ≅1.618 aka ɸ aka ‘the golden ratio’.  It’s found throughout history, where it has influenced architects and artists. We see it in nature, in plants, and it is used in movies to frame shots. It can be approximated by the Fibonacci sequence (another math favorite of mine). However, I have not worked out how to apply this to flow cytometry.  That doesn’t mean numbers aren’t important in flow cytometry. They are central to everything we do, and in this blog, I’m going to flit around numbers-based questions that I have received…

How To Do Variant Calling From RNASeq NGS Data

How To Do Variant Calling From RNASeq NGS Data

By: Deepak Kumar, PhD

Developing variant calling and analysis pipelines for NGS sequenced data have become a norm in clinical labs. These pipelines include a strategic integration of several tools and techniques to identify molecular and structural variants. That eventually helps in the apt variant annotation and interpretation. This blog will delve into the concepts and intricacies of developing a “variant calling” pipeline using GATK. “Variant calling” can also be performed using tools other than GATK, such as FREEBAYES and SAMTOOLS.  In this blog, I will walk you through variant calling methods on Illumina germline RNASeq data. In the steps, wherever required, I will…

Understanding Clinical Trials And Drug Development As A Research Scientist

Understanding Clinical Trials And Drug Development As A Research Scientist

By: Deepak Kumar, PhD

Clinical trials are studies designed to test the novel methods of diagnosing and treating health conditions – by observing the outcomes of human subjects under experimental conditions.  These are interventional studies that are performed under stringent clinical laboratory settings. Contrariwise, non-interventional studies are performed outside the clinical trial settings that provide researchers an opportunity to monitor the effect of drugs in real-life situations. Non-interventional trials are also termed observational studies as they include post-marketing surveillance studies (PMS) and post-authorization safety studies (PASS). Clinical trials are preferred for testing newly developed drugs since interventional studies are conducted in a highly monitored…

How To Profile DNA And RNA Expression Using Next Generation Sequencing (Part-2)

How To Profile DNA And RNA Expression Using Next Generation Sequencing (Part-2)

By: Deepak Kumar, PhD

In the first blog of this series, we explored the power of sequencing the genome at various levels. We also dealt with how the characterization of the RNA expression levels helps us to understand the changes at the genome level. These changes impact the downstream expression of the target genes. In this blog, we will explore how NGS sequencing can help us comprehend DNA modification that affect the expression pattern of the given genes (epigenetic profiling) as well as characterizing the DNA-protein interactions that allow for the identification of genes that may be regulated by a given protein.  DNA Methylation Profiling…

How To Profile DNA And RNA Expression Using Next Generation Sequencing

How To Profile DNA And RNA Expression Using Next Generation Sequencing

By: Deepak Kumar, PhD

Why is Next Generation Sequencing so powerful to explore and answer both clinical and research questions. With the ability to sequence whole genomes, identifying novel changes between individuals, to exploring what RNA sequences are being expressed, or to examine DNA modifications and protein-DNA interactions occurring that can help researchers better understand the complex regulation of transcription. This, in turn, allows them to characterize changes during different disease states, which can suggest a way to treat said disease.  Over the next two blogs, I will highlight these different methods along with illustrating how these can help clinical diagnostics as well as…

What Is Next Generation Sequencing (NGS) And How Is It Used In Drug Development

What Is Next Generation Sequencing (NGS) And How Is It Used In Drug Development

By: Deepak Kumar, PhD

NGS methodologies have been used to produce high-throughput sequence data. These data with appropriate computational analyses facilitate variant identification and prove to be extremely valuable in pharmaceutical industries and clinical practice for developing drug molecules inhibiting disease progression. Thus, by providing a comprehensive profile of an individual’s variome — particularly that of clinical relevance consisting of pathogenic variants — NGS helps in determining new disease genes. The information thus obtained on genetic variations and the target disease genes can be used by the Pharma companies to develop drugs impeding these variants and their disease-causing effect. However simple this may allude…

7 Key Image Analysis Terms For New Microscopist

7 Key Image Analysis Terms For New Microscopist

By: Heather Brown-Harding, PhD

As scientists, we need to perform image analysis after we’ve acquired images in the microscope, otherwise, we have just a pretty picture and not data. The vocabulary for image processing and analysis can be a little intimidating to those new to the field. Therefore, in this blog, I’m going to break down 7 terms that are key when post-processing of images. 1. RGB Image Images acquired during microscopy can be grouped into two main categories. Either monochrome (that can be multichannel) or “RGB.” RGB stands for red, green, blue – the primary colors of light. The cameras in our phones…

We Tested 5 Major Flow Cytometry SPADE Programs for Speed - Here Are The Results

We Tested 5 Major Flow Cytometry SPADE Programs for Speed - Here Are The Results

By: Tim Bushnell, PhD

In the flow cytometry community, SPADE (Spanning-tree Progression Analysis of Density-normalized Events) is a favored algorithm for dealing with highly multidimensional or otherwise complex datasets. Like tSNE, SPADE extracts information across events in your data unsupervised and presents the result in a unique visual format. Given the growing popularity of this kind of algorithm for dealing with complex datasets, we decided to test the SPADE algorithm in 5 software packages, including Cytobank, FCS Express, FlowJo, R, and the original, free software made available by the author of SPADE. Which was the fastest?

5 FlowJo Hacks To Boost The Quality Of Your Flow Cytometry Analysis

5 FlowJo Hacks To Boost The Quality Of Your Flow Cytometry Analysis

By: Tim Bushnell, PhD

FlowJo is a powerful tool for performing and analyzing flow cytometry experiments, if you know how to use it to the fullest. This includes understanding embedding and using keywords, the FlowJo compensation wizard, spillover spreading matrix, FlowJo and R, and creating tables in FlowJo. Extending your use of FJ using these hacks will help organize your data, improve analysis and make your exported data easier to understand and explain to others. Take a few moments and explore all you can do with FJ beyond just gating populations.

Top Industry Career eBooks

Get the Advanced Microscopy eBook

Get the Advanced Microscopy eBook

Heather Brown-Harding, PhD

Learn the best practices and advanced techniques across the diverse fields of microscopy, including instrumentation, experimental setup, image analysis, figure preparation, and more.

Get The Free Modern Flow Cytometry eBook

Get The Free Modern Flow Cytometry eBook

Tim Bushnell, PhD

Learn the best practices of flow cytometry experimentation, data analysis, figure preparation, antibody panel design, instrumentation and more.

Get The Free 4-10 Compensation eBook

Get The Free 4-10 Compensation eBook

Tim Bushnell, PhD

Advanced 4-10 Color Compensation, Learn strategies for designing advanced antibody compensation panels and how to use your compensation matrix to analyze your experimental data.